CJ1W-OC211 Contact Output Unit (16 Points) Name 16-point Contact Output Unit with Terminal Block Model CJ1W-OC211 Max. Switching Capacity 2 A 250 VAC (cosφ = 1), 2 A 250 VAC (cosφ = 0.4), 2 A 24 VDC (8 A/Unit) Min. Switching Capacity 1 mA 5 VDC Relays NY-24W-K-IE (Fujitsu Takamizawa Components, Ltd.), Cannot be replaced. Service Life of Relay Electrical: 150,000 operations (24 VDC, resistive load)/ 100,000 operations (250 VAC, cosφ = 0.4, inductive load) Mechanical: 20,000,000 operations Service life will vary depending on the connected load. ON Response Time 15 ms max. OFF Response Time 15 ms max. Number of Circuits 16 points/common, 1 circuit Insulation Resistance 20 MΩ between external terminals and the GR terminal (500 VDC) Dielectric Strength 2,000 VAC between the external terminals and the GR terminal for 1 minute at a leakage current of 10 mA max. Internal Current Consumption 110 mA 5 VDC max. 96 mA 24 VDC max. (6 mA × No. of ON points) Weight 170 g max. Circuit Configuration The signal names of the terminals are the device variable names. The device variable names are the names that use ''Jxx'' as the device name. External connection and terminal-device variable diagram The signal names of the terminals are the device variable names. The device variable names are the names that use ''Jxx'' as the device name. * Terminal numbers A0 to A8 and B0 to B8 are used in the external connection and terminal-device variable diagrams. They are not printed on the Units.
read more